화학공학소재연구정보센터
Chemical Engineering Science, Vol.66, No.16, 3701-3709, 2011
Porometry, porosimetry, image analysis and void network modelling in the study of the pore-level properties of filters
We present fundamental and quantitative comparisons between the techniques of porometry (or flow permporometry), porosimetry, image analysis and void network modelling for seven types of filter, chosen to encompass the range of simple to complex void structure. They were metal, cellulose and glass fibre macro- and meso-porous filters of various types. The comparisons allow a general reappraisal of the limitations of each technique for measuring void structures. Porometry is shown to give unrealistically narrow void size distributions, but the correct filtration characteristic when calibrated. Shielded mercury porosimetry can give the quaternary (sample-level anisotropic) characteristics of the void structure. The first derivative of a mercury porosimetry intrusion curve is shown to under estimate the large number of voids, but this error can be largely corrected by the use of a void network model. The model was also used to simulate the full filtration characteristic of each sample, which agreed with the manufacturer's filtration ratings. The model was validated through its correct a priori simulation of absolute gas permeabilities for track etch, cellulose nitrate and sintered powder filters. (C) 2011 Elsevier Ltd. All rights reserved.