화학공학소재연구정보센터
Electrochimica Acta, Vol.271, 35-40, 2018
Electrochemical measurement of lamotrigine using boron-doped diamond electrodes
In this paper we report on a study of the electrochemical behavior of lamotrigine (LTG) using boron-doped diamond (BDD) electrodes. Cyclic voltammetry of LTG in a phosphate buffer solution exhibited an oxidation signal at ca. +1.4 V (vs. Ag/AgCl). A reduction signal at -0.5 to -1.0 V (vs. Ag/AgCl) was observed only after oxidation of LTG. The oxidation signal was due to polymerization of LTG adsorbed on the BDD electrode, while the reduction signal was a consequence of a process involving the reduction of the azo group in the LTG-oxidized product. Continuous measurements of LTG aiming at in vivo monitoring were successfully investigated and an optimized chronoamperometry technique, in which the BDD microelectrode was clamped to the LTG oxidation potential (+1.4 V) for 2 s, then to the LTG reduction potential (-0.6 V) for 2 s, was implemented. The limits of detection (S/N = 3) were estimated to be 29 nM and 130 nM for the oxidation and reduction of LTG, respectively. The use of this chronoamperometry technique enables a high temporal resolution of 4 s for LTG measurements. BDD microelectrodes, therefore, are the ideal electrodes for performing continuous in vivo measurements of LTG. (C) 2018 Elsevier Ltd. All rights reserved.